|
Equipment Leica SP5 confocal Zeiss 510 inverted Zeiss 510 upright Two photon Spinning disk TIRF Axio Observer Axio Imager Live cell station Stereoscope Microinjection Gizmos Image analysis Services New Users Server Booking Rules Fees About LMCF Personnel Contacts Publications Image Gallery Reference Guides FAQs Feedback Ask a question Report a problem Bugs Search |
Zeiss LSM 410 confocal manual Welcome The Zeiss 410 is the Comprehensive Cancer Center's shared use confocal, which is maintained and supported by the Light Microscopy Core Facility. The instrument is a robust imaging system based on the Zeiss Axiovert 100 inverted microscope. UV, green, red and far-red laser lines are available and high-quality images of up to three fluorophores can be produced simultaneously. LMCF staff are here to help you - we will train you to use the instrument and will provide technical support and imaging assistance as needed. If you have any queries or difficulties, please contact us at LMCF@duke.edu or 613-8168. Conventions used in this manual
Words of precaution Using oil immersion objectives
If you fail to follow these instructions, there is significant risk of oil seeping down into the objectives and causing expensive damage to the optics of the microscope!! Risk of damage when focusing on specimenIf you ever feel ANY resistance when turning the focus knob (whether going up or down) STOP TURNING THE FOCUS IMMEDIATELY! Because of the mechanical advantage in the coarse adjust, it takes very little force to damage the objectives and/or the focus control gearing. The top of 40x or 63x objectives should be even with the top of the stage plate when those objectives are in focus. Set this first by eye before you put your slide on the stage, then change to the low power objective that you wish to use, place your slide on the stage, and adjust focus using ONLY the fine focus control If you follow this simple procedure, you won't damage to the microscope, the objectives, or your specimen. More complicated preparations (like multiwell plates) require more complicated procedures and greater care on the part of the user. PLEASE see Heather or Sam if you plan on using an unusual setup. System Start Up
Conventional Microscopy Put slide on stage oriented perpendicular to front edge of table with its coverslip down as in Figure 1. The coverslip should fit entirely within the hole in the center of the stage (use a 22x30 mm or smaller coverslip). You may add a small piece of tape at one end of the slide to secure it, but this usually is not necessary. Most objectives are designed to use #1.5 coverslips. Using the wrong one may have serious implications on image quality. Please refer to the link for "coverslip thickness" which can be found on the LMCF website.
Fig 1) Proper Orientation Of Slide On Stage
Fig 2) Diagram of Confocal Mirror and Fluorescence Filters In Reflector Slider If you have trouble finding your specimen, close the condenser diaphragm all the way (small lever on front of condenser all the way to the right). This will improve contrast and make your specimen easier to find.
Fig 3) Condenser and Floating Stage Avoid actually touching the objective with the applicator. Avoid getting bubbles in the oil by letting the oil flow into the nozzle by gravity for a moment before squeezing the bottle very gently. Continue rotating the nosepiece until the objective clicks into position beneath the slide. Reminder: once you have used an oil immersion objective on a slide you cannot use the 10 or 20X dry objectives with that slide. In fact, you must use caution when changing objectives to avoid contaminating the dry objectives.
Fig 5) Fluorescence Shutter Closed & Open Single channel confocal image acquisition 1. Close the fluorescence shutter (slide fully to the right). Move the reflector slider all the way to the right to the "confocal mirror" position (you may hear a very short beep). 2. Click the toggle button <> between Conv and LSM in the "Control Panel". 3. Open the fluorescence shutter (slide over one position to the left). 4. Check the red-green laser power output by clicking POWER above the "Control Panel" near top of screen. The power level should be set to 30 for most applications. If using UV laser, set its power level according to the UV instructions near the end of this manual. 5.Under the Lens= menu at the top of the "Control Panel", select the objective you are using. The Lens setting is very important! Since there is no electronic communication between the computer and the nosepiece, all calculations made by the software (scale bars, z projections, etc) will be derived from this information you provide. 6. Under the At= (attenuation) menu, you will find 3 columns of numbers (1, 10, 30, etc). This menu selects neutral density filters for the laser light entering the microscope. The left column controls the far-red laser (for Cy5, Alexa633), the middle column controls attenuation of the UV laser, and the right column controls the red/green laser. Start with the far-red and red/green lasers at 10 (=1/10 laser light) and the UV laser at 3. If you find you need more excitation power, then set the attenuation to a lower number. 7. Select the proper dichroic mirror by putting the lever on the front panel of the scanning unit marked DBS2 in the "Free" position. This is as per Table 1 (the "Cheat Sheet" describing the dichroic positions for single, double, and triple channel imaging combinations) which is included in this manual and posted in the imaging suite. ![]() Table 1) Dichroic Mirrors and Emission Filters Cheat Sheet 8. For green-labeled specimens, click radio button 1 in the Setup box on left side of the "Control Panel". For red imaging, click 2. For far-red, click 3. 9. In the Scan box, click Start. You should see a flickering light at the tip of the objective. If not, check the position of the fluorescence shutter (step 3). 10. Adjust focus with the fine focus knob until you see the specimen on the display monitor. You may see red & blue pixels in the image - this is the range indicator color table (Image→ Color Tables→ Range Indicator if it appears you are not in this mode). Range indicator is used in setting brightness & contrast of the image. Adjust focus until you see the brightest image, which will probably have many red pixels in the areas where signal is detected. 11. Press F9 on the keyboard to obtain initial brightness (offset) and contrast (gain) settings. Optimize these settings by adjusting the Contrast scroll bar (click and drag slider with the right mouse button for fine adjust) so that about 10% of the pixels are red (these will be the brightest ones in the final image). Adjust the Brightness scroll bar so that the background is no more than 50% speckled with blue pixels where you do not expect signal (these will be the darkest pixels in your final image). 12. Recheck focus. If you find that the image becomes much brighter, then readjust contrast and brightness as before according to this brighter focal plane. Click Stop when not actively adjusting the image to avoid bleaching your specimen. 13. To improve image quality, select "Averag" in the Extras box of the "Control Panel". Select line averaging and the number of iterations you would like the laser to scan each line. Click Done in the Averaging window. The amount of averaging necessary to obtain a good image depends on the signal to noise ratio of your sample image. Low contrast images will require more averaging but be aware that this could cause photobleaching. You can determine the amount of averaging necessary to yield the best image of your sample empirically, but most samples will require averaging of 4 or 8. 14. To collect the averaged image, click Single in the Scan box. 15. For a gray scale display, click Color in the "Display Control" window 3 times or select Image→ Color Tables→ gray scale from the main menu. For a colorized display, click the RGB radio button in the Image box of the "Display Control". 16. To save the image, choose File→ Store Image. You should see the contents of the Bio-lemming\Confocal server, which has been mapped to the H: drive. (The FROG confocal server is mapped to the F: drive.) Scroll down to your lab's folder, then double-click to open. Repeat as necessary to enter your personal user folder. Clear the contents of the File box and type a name (up to 8 characters) using lower case letters and/or numbers only. Click Store to save image. Files are saved as a Tiff, which can be read on either PC or Macintosh computers. You can use Windows to make new folders with longer names describing your experiment - ask us if you do not know how to do this. 17. To increase magnification of a scan:
18. To switch back to conventional microscopy:
Dual or triple channel confocal image acquisition
Because of the 8 character limit to the length of file names, the pre-set parameters for general use are named after the cyanin fluorochromes with UV fluors referred to as "dapi" or "blu". Therefore, it is important to know that Cy2 is a green fluor, Cy3 is a red fluor, Cy5 is a far-red fluor. Considering this naming convention, "-cy2cy3" is the setting for double-channel red/green imaging and "-c2c3blu" is the triple-channel setting for imaging the green, red, and UV fluor combination. These are all explained in the cheat sheet. Double- or triple-channel imaging using the SEQUEN function In some cases (e.g. severe bleedthrough conditions) it is better to sequentially acquire various color channels of data from a sample. This can be done manually, by exciting each fluorochrome individually and saving the images to different color channels of the video monitor. The various color channels are overlaid by clicking RGB in the "Display Control", and the composite image can be saved in the usual manner (File→ Store Image). The SEQUEN and SEQUEN3 functions automate the sequential excitation, acquisition, and display overlay of this process. Their use is described here.
Fig 6) Location of SEQUEN Buttons In Software Z-sectioning 1. After setting the brightness and contrast for every channel in your specimen, determine the start and end points of the Z scan. Z sectioning is always done moving into the specimen, away from the coverslip. To find the starting point for the Z series, Start scanning (with the averaging off) and focus toward the coverslip (rotate fine focus knob toward you) until you reach the point at which you'd like to begin the Z series. Stop scanning. 2. Select Z→ Initialize motor from the main menu. The Z value in the "Parameters" window should now read 0.0. 3. Start scanning. Turn fine focus knob away from you (counterclockwise) focusing down through your specimen until you reach the point at which you'd like to end the Z series. Stop scanning and note the Z value at this point in the specimen. 4. Manually focus back to the Z=0.0 position. Set averaging to your desired value. 5. Select Z→ Z Sectioning under the main menu. Set your desired Z interval (optimal settings depend on the pinhole size, the objective and the wavelength of the light but as a starting point try 1 µm for 40x objective, 0.5 µm for 63x objective). Set the number of sections to cover the interval between the start and finish points of the series:
Set current section position to 1. 6. Specify where the Z series files will be stored. Select File in the Destination box. Click Dir that appears to the right to select the folder to which the series will be saved. Select your folder on the server (H: drive). Type in an alphanumeric file name up to 6 characters long (do not end with a number). Click Select. The name you provided should now appear in the box just to the left of Dir. 7. Click OK to begin Z sectioning. Each image of the series will be stored as an individual tiff file with numbers (starting with 00) appended to the name you specified. Z-Sectioning using the DUALSCAN macroThe DUALSCAN macro sequentially excites & acquires multiple fluorochromes in a single specimen at fixed intervals over a specified distance. Currently, this macro is set up to collect only two color channels of information from a sample. DUALSCAN works by exciting the first fluorochrome, collecting its emission signal, and writing that data to a designated color channel in video memory. The macro then excites the second fluorochrome, collects its emission, and writes it to a contrasting color channel of video memory. Finally, it displays the overlay of the two collected images before moving on to the next user-defined z-interval. The user can specify to save the monochrome image series of each color channel and/or that of the RGB overlaid images. Sequentially excited images are saved to the directory as they are acquired.
Fig 7) Dualscan Window number of sections needed = (thickness of sample / z interval) + 1
Fig 8) Z Sectioning Window Fig 9) Directory Path Selection Window Pressing Break will stop the dualscan image acquisition, and the focus drive (z motor) returns to the top of the z Images taken to that point in the acquisition will have been saved on the specified directory path. Ending your session 1. Sign out on the logbook by listing end time. Indicate whether or not the system was left on for the next user. Note any problems in the Comments column of the logbook. 2. Select File→ Quit from the main menu or close the program using the X in the upper right hand corner of the main window. 3. 3. Check the on-line reservation schedule. (Use the Internet Explorer icon on the 410’s desktop.) If someone else is scheduled to use the microscope within 2 hours, then select the "Log Off" command from the Start menu of the desktop and leave everything running. The log in window should appear. However, if it is near the end of the day or on the weekend, please contact the next person on the schedule to ensure that they are actually planning to use the microscope. If you can't contact them, then shut off the system. 4. If you have been using the UV laser and no one else is signed up to use the UV in the next 2 hours, turn it off now. Leave the chiller running to cool the UV laser down for the remainder of the shut down process (at least 3 minutes). Refer to the "Using the UV laser" for full details of UV laser power down. To turn off the system, select "Shut down computer" command in the Start menu of the desktop. Wait for a message stating "It is now safe to turn off the computer" to appear. Turn off the computer by turning the key on the cabinet below the air table to the off position. 6. Turn the key switch on the KrAr laser (on shelf above the microscope) to the off position. The fan will continue to run for several minutes then will shut off automatically. 7. Turn off the mercury lamp (on shelf above the microscope). 8. To make cleaning the microscope stage and objectives easier, gently tilt the upright condenser stand back out of the way. 9. Clean oil immersion objectives by gently removing the excess oil from the objective surface with lens paper and/or a cotton-tipped swab. Do not use any solvent! The non-immersion objectives (5x, 10x, 20x) should not be cleaned by users of the facility. If you note oil on a non-immersion objective then notify LMCF at lmcf@duke.edu. 10. Rotate the objective holder by its knurled edge (not by objective) to put the 5x or 10x objective in place. 11. If the stage and/or benchtop surfaces have oil on them, wipe them clean using a kimwipe dampened with alcohol. 12. Return the condenser stand to vertical position. 13. Cover the microscope with the blue cover. If the mercury lamphouse is still hot, cover everything except the lamphouse. 14. Turn off the UV chiller switch if necessary. 15. Turn off the room lights and close the door behind you (unlocked). Make sure the main door to the Imaging Suite (4226) is locked behind you when you leave. This door should always be kept locked when nobody is there. 16. When you return to your lab, download your files onto some other storage media (e.g. your department's fileserver, memory stick, CD/DVD) right away. If you do not know how to do this, then please seek help from your department’s computer support personnel. Files will remain on the server for 1 week, but after that time are subject to deletion without notice. Using the UV laser General guidelines for UV usage
Turning the UV laser ON/OFF
Preparing to image with the UV laser
The UV laser requires a 20-30 minute warm-up period for optimal excitation of your specimen. Hints for better UV imaging
Image sequence applications All of these features require that a sequence of files be preloaded into the host memory (or RAM) of the computer.
Animate
Gallery
The projection software creates 3-dimensional views using alpha blending as a model for reconstruction. It is easiest to think of alpha blending as illuminating the specimen from different angles or viewpoints. A maximal projection is one which takes the intensity values from all sections and collapses them into one single illuminated image. "Maximum" overlay is selected when the maximum pixel intensity is desired to be represented in the projection. "Transparency" overlay mode decreases the brightness of each file in the projected stack to reveal structure in subsequent layers that would otherwise be hidden. Creating a maximal projection is described here.
The Zeiss software can provide scale bars for any confocal image. For these to be accurate however, you must have selected the correct objective under the lens menu (see step 5 under the "Single channel confocal image acquisition") before you collected and saved the image. Normally the scale bar is written into a channel called the overlay channel. Although it appears on the screen, it is not written into a standard part of a tiff file and therefore does not appear on the screen when a file is exported into another program such as Photoshop. There are two methods for transferring the bar from the overlay channel into one of the regular channels so that it can be exported. Either will work on the confocal microscope computer itself, while only the latter will work on the workstation computer. Method I (Confocal microscope computer only) This method uses a macro called OVLTORGB (overlay to RGB), which is located near the top of the Zeiss confocal software menu. This macro converts the normal scale bar (held in a part of memory called the overlay channel) into a line that is written onto the image. To produce a scale bar:
Because the scale bar has jagged edges, you will probably want to save this modified file under a different name - once the OVLTORGB command is performed you cannot remove the scale bar from the image. You can then use this file as a guide for making a new neater looking scale bar on the original image using Photoshop, Powerpoint or another program. Method II (Confocal microscope or workstation computer) Method II uses the measure function to provide crosshairs on the image at a defined distance that can then be used to produce a scale bar. This method will work on either computer.
|